Product/Service

MPPI-2 Pressure Injector

Source: Applied Scientific Instrumentation
MPPI-2 Pressure Injector
  • Front panel push button switch
  • External TTL input signal
  • Optional footswitch

    Linear adjustment of both pulse width and pressure gives precise control to meet all of your pressure injection needs.

    The MPPI-2 utilizes an external filter for reliability and ease of maintenance. The input pressure can be up to 100psi. The output pressure is adjustable from 0 up to 100psi.

    The MPPI-2 has an input for an optional Back Pressure Unit. This option precisely regulates leakage from the injection pipette. The BPU provides a convenient means of providing an adjustable, 0 to 15 psi, back pressure to the MPPI-2. The back pressure is applied to the injection pipette between injections to prevent back filling. Without the BPU the pipetter is at atmospheric pressure between injections.

    The Milli-Pulse Pressure Injector is a self contained device for producing gas pressure pulses to an injection pipette. The unit offers linear control of both pressure and pulse duration. The unit provides three convenient means of initiating the pressure pulse:

    Specifications

    Input Pressure Range	-	0 to 300 p.s.i.
    
    Output Pressure Range	-	Regulated 0 to 100 p.s.i.(Maximum output is 									dependent upon input pressure )
    
    Output Pulse Range	-	Three factory set ranges:(please specify when ordering)
    				-1 to 100ms
    				-0 to 1 second
    				-0 to 10 seconds
    Minimum duration is limited by the pneumatic solenoid valve to approximately 5ms
    
    Output Pulse Accuracy		-	0.4% of full scale (crystal controlled)
    
    Output Pulse Repeatability	-	0.4% of full scale (1% over the life of the valve)
    
    Valve Life Expectancy		-	100 million cycles
    
    Output Pulse Setting		-	Via a front panel 10 turn calibrated dial
    
    Modes of Operation		-	-Pulsed pressure via pushbutton, 
    					 external TTL pulse or optional footswitch
    					-Continuous pressure 
    					-Gated pressure via external TTL level input
    
    Pulsed Pressure Modes		-	-Front panel pushbutton switch
    					-External TTL input signal
    					-Optional footswitch
    
    Output Pressure Gauge		-	Front panel analog gauge 0 to 100 psi
    
    Power Requirements		-	120v AC, 60 Hz, 30 watt or 220v, 60 Hz
    					Please specify when ordering
    						
    Gas Input & Back 		-	1/8" barbed hose fittings
    Pressure Fitting		
    	
    Gas Output Fitting		-	1/16" barbed hose fitting
    
    Recommended Gas			-	Nitrogen or clean dry compressed air, an 
    					externally mounted input gas filter is provided
    
    Size				-	Compact bench top unit measuring

    Applied Scientific Instrumentation, 29391 West Enid Road, Eugene, OR 97402-9533. Tel: 541-461-8181; Fax: 541-461-4018.